SmartReader Data Loggers
88
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
(100ft) leads (which may add an extra 10 ohm resistance) an error of less than 0.025°C will be
introduced. The higher the resistance of the thermistor, the more negligible this error will
become.
Keeping this in mind you will be able to run cables up to 30m (100ft) without appreciable
accuracy degradation. If you intend on using longer cables, such factors as electrical noise
and capacitance must be considered.